8
PhenoMaster | Hardware Operating Instructions
2.
Instructions for safe Operation
See Hardware Operating Instructions - Core Modules, Section "Instructions for safe Operation".
2.1. General Safety Instructions
Proper Use
2.1.1.
NOTICE
- Reduced lifetime!
If Calorimetry Process Control Units and Air Pumps are powered ON without operating the
system, this may reduce lifetime of devices.
•
Always run system software when devices are powered ON.
•
Power OFF devices after use.
NOTICE
– Reduced lifetime!
If High-Speed Sensor Unit is powered ON without operating the system, contamination may
damage the sensors.
•
During operating mode of High-Speed Sensor Unit, ensure that…
o
…its sensors are supplied with air from reference gas cylinder continuously.
o
…all other devices of the CaloSys module entered operating mode as well.
o
…PhenoMaster software is running.
•
Power OFF High-Speed Sensor Unit after use.
NOTICE
- Reduced lifetime!
If excess heat produced by the Air Drying Units cannot dissipate to the outside, this may
reduce lifetime.
Ensure that the device is properly ventilated:
•
Min. distance:
o
Air Drying Unit - Wall: >=150…200mm (>=5.91…7.87inch)
o
Air Drying Unit - Other devices: >=150…200mm (>=5.91…7.87inch).
NOTICE
– Reduced lifetime!
If Calorimetry Process Control Units and Air Pumps are powered ON while there is a leakage
in the sample line, this may reduce lifetime of devices.
•
After you have discarded condensed water, carefully reconnect collection vessel to 1x Air
Drying Unit and reconnect separating bottle to tubing.
•
Select "Status/Calo" and check for proper "SampleFlow” on "Calo Current Values" dialog (see
Software Operating Instructions – Calorimetry, Section "Status Menu, Calo").
NOTICE
– Reduced lifetime!
If Calorimetry Process Control Units and Air Pumps are powered ON while there is a leakage
in the sample line, this may reduce lifetime of devices.
•
Never remove collection vessels from the 4x Air Drying Unit. Drain off condensed water via
collection vessel tubing only.
•
After you have discarded condensed water of separating bottle, carefully reconnect bottle to
tubing.
•
Select "Status/Calo" and check for proper "SampleFlow” on "Calo Current Values" dialog (see
Software Operating Instructions – Calorimetry, Section "Status Menu, Calo").
Содержание Calorimetry PhenoMaster
Страница 6: ...6 PhenoMaster Hardware Operating Instructions info TSE Systems com www TSE Systems com ...
Страница 13: ...CaloSys Modules 13 Confirm default installation path by pressing Next info TSE Systems com www TSE Systems com ...
Страница 15: ...CaloSys Modules 15 Press Finish info TSE Systems com www TSE Systems com ...
Страница 69: ...CaloSys Modules 69 info TSE Systems com www TSE Systems com ...