CaloSys Modules
39
Connections CaloSys Module - Pull Principle Respirometry - Automatic or Manual Calibration
Port
Function
Vacuum Pump
"Vac Inlet"
Tube connection to Calorimetry Master Control Unit via pressure control valve.
Calorimetry Master Control Unit
"X01"
37-pin connection to PC.
"X02"
37-pin connection to PC.
"X03"
37-pin connection to PC.
"X04"
25-pin connection to Sample Switch Unit.
"X05"
37-pin connection to Slave Control Unit - if available.
"X07"
37-pin connection to Slave Control Unit - if available.
"X08"
9-pin connection to PC via USB-RS485 converter box - if available. Allows to switch between MFC mouse
(1l/min) and MFC rat (5l/min).
"X0 ext"
Connection to Calibration Control Unit.
"Flow IN 1...8"
Tube connections to animal cages.
"Flow OUT"
Tube connection to vacuum pump.
"Sample1 IN"
Tube connection Calibration Unit.
"Sample1 OUT"
Sample air outlet.
"Ref. Air"
Tube connection to reference cage.
"Temperature 1...8"
5-pin connection to temperature sensors.
Animal Cage
"Air OUT"
Tube connection to Calorimetry Master Control Unit, used air.
"Air IN"
Inlet, Fresh air supply.
"Sample"
Tube connection to Sample Switch Unit.
Sample Switch Unit
"1...8"
Tube connections for animal cage 1...8.
"Ref. IN"
Tube connection for reference air.
Sample "OUT"
Tube connection to Air Drying Unit.
"X04"
25-pin connection to Calorimetry Master Control Unit.
Air Drying Unit
Sample "IN"
Tube connection to sample switch pump.
Sample "OUT"
Tube connection to Calibration Control Unit via separating bottle.
Calibration Control Unit
"X0 ext"
Connection to Calibration Control Unit.
"Sample IN"
Tube connection to Air Drying Unit via separating bottle.
"Sample OUT"
Tube connection to Calorimetry Master Control Unit.
"Gas Inlets1...4"
Tube connections to calibration gas cylinders via pressure regulator (automatic calibration).
PC
"X01" PCI 9111
37-pin cable connection to Calorimetry Master Control Unit.
"X02" PCI 9111
37-pin cable connection to Calorimetry Master Control Unit.
"X03" PCI 9114
37-pin cable connection to Calorimetry Master Control Unit.
"USB"
Connection to High-Speed Sensor Unit via USB-RS485 converter.
USB
Connection to Calorimetry Master Control Unit via USB-RS485 converter box - if available. Allows to switch
between MFC mouse (1l/min) and MFC rat (5l/min) (not shown).
High-Speed Sensor Unit
Separate Sensor Unit for fast O2 and CO2 measurements. For further instructions, see Hardware Operating
Instructions - High-Speed Sensor Unit.
In systems with a multiple of eight measuring places, further Slave Units are included in the system to
connect up the additional calorimetry cages.
Содержание Calorimetry PhenoMaster
Страница 6: ...6 PhenoMaster Hardware Operating Instructions info TSE Systems com www TSE Systems com ...
Страница 13: ...CaloSys Modules 13 Confirm default installation path by pressing Next info TSE Systems com www TSE Systems com ...
Страница 15: ...CaloSys Modules 15 Press Finish info TSE Systems com www TSE Systems com ...
Страница 69: ...CaloSys Modules 69 info TSE Systems com www TSE Systems com ...