TH2826 Operation manual
Ver1.0
◇
132
30
/INDEX
SEQ
:
When the analog test of the last sweep point is finished and
the UNKNOWN terminal can be connected to another DUT,
/INDEX is effective. But the comparison signal is effective until
/EOM is effective.
STEP
:
When the analog test of the last sweep point is finished,
/INDEX is effective. But the comparison signal is effective until
/EOM is effective.
31
/EOM
Test ended
:
SEQ
:
when the test is finished and the compared results are
effective, this signal is effective.
STEP
:
When the test of every sweep point is finished, /EOM is finished.
the comparison result signal is effective until /EOM is effecive.
others
The definition is the same to that of the comparison.
/BIN1
/BIN2 /BIN3 /BIN4 /BIN5 /BIN6 /BIN7 /BIN8 /BIN9 /OUT
the measurement data
the high limit
the good area
the low limit
1 2 3 4 5 6 7 8 9 10
the sweep point
Figure 9-4 the signal area of the list sweep comparison