TH2826 Operation manual
Ver1.0
◇
29
Open full clear
DCR open
2)
Connect fixture to test terminal.
3)
Press
Open full clear
,
TH2826 can test open admittance of 41frequency points. It costs
75 seconds for open correction. The following softkeys are displayed.
Drop
The softkey can cancel the current open correction. The former data is reserved.
4)
Press
DCR open
,
TH2826 can test open resistance.
5)
Press
ON
, the open correction is valid
,
TH2826 then perform open correction calculation .
If frequency1
,
frequency 2 and frequency 3 is set as OFF, the insert method is adopted to
calculate the date. If frequency1
,
frequency 2 and frequency 3 is set as ON, meanwhile
the test frequency is equal to frequency1
,
frequency 2 or frequency 3
,
then open data of
frequency1
,
frequency 2 and frequency 3 is used to calculate the open correction.
6)
Press
OFF
,
turn off the correction function.
3.5.2
Short correction
The open correction of TH282 can clear the error caused by parasitic admittance.
(
R,
X
)
,shown as figure 3-8.
Figure 3-8
TH2826 adopts two open corrections for data.
TH2826 can correct the following 41 fixed frequency points. Except the following
points
,
the instrument can calculate the open correction data. Move cursor to
Short
,
use
Short full clear
to execute the short correction. The 41 fixed frequency points is
the same as open correction.
TH2826 can set 3 open correction frequency points
:
Freq 1
,
Freq 2
or
Freq 3
Move
cursor to
Freq 1
,
Freq 2
or
Freq 3.
Use
Short single clear
to correct 3 set
frequency.
Содержание TH2826/A
Страница 49: ...TH2826 Operation manual Ver1 0 41 Figure 3 15 ...