TH2826 Operation manual
Ver1.0
◇
28
The test result of load correction.
(
Measure A
,
Measure B
)
In multi way correction mode
,
the current channel number(
Channel
)
The actual result of load correction can be tested on
Freq 1
,
Freq 2
and
Freq 3
zone.
3.5.1
Open correction
The open correction of TH282 can clear the error caused by scattering admittance. Shown as
figure 3-7.
Figure 3-7
TH2826 adopts two open corrections for data.
TH2826 can correct the following 55 fixed frequency points. Except the following
points
,
the instrument can calculate the open correction data. Move cursor to
Open
,
use
Open full clear
to execute the open correction. Shown as the figure 3-1.
Figure 3-1
20
25
30
40
50
60
80
[Hz]
100
120
150
200
250
300
400
500
600
800
[Hz]
1
1.2
1.5
2
2.5
3
4
5
6
8
[kHz]
10
12
15
20
25
30
40
50
60
80
[kHz]
100
120
150
200
250
300
400
500
600
800
[kHz]
1
1.2
1.5
2
2.5
3
4
5
[MHz]
TH2826 can set 3 open correction frequency point
:
Freq 1
,
Freq 2 and Freq 3. Move
cursor to
Freq 1
,
Freq 2
or
Freq 3
,
use
open single clear
to correct 3 set
frequency.
Operational step of open correction
1)
Move cursor to
Open
,
the following softkeys are displayed.
ON
OFF