TH2826 Operation manual
Ver1.0
◇
111
length of the cable.
command syntax
:
CORRection:LENGth <value>
<value>:0, 1 or 2 with the unit m as the postfix.
e.g.
:
WrtCmd(“CORR
:
LENG 1m”) :set the cable length to 1m.
query syntax
:
CORRection:LENGth?
return
:
<NR1><NL^END>
:METHod
is used to set the correction mode. Char ‘?’ can be used to query the correction
mode. command syntax
:
CORRection:METHod
MULTi
SINGle
SINGle:set or return the single channel mode.
MULTi: set or return the multi-channel mode.
e.g.
:
WrtCmd(“CORR:METH MULT”) ;set the instrument to multi-channel mode
query syntax
:
CORRection:METHod?
return
:
SINGle
<NL^END>
MULTi
:OPEN
is used to perform the correction data of 41 preset test points.
command sytax
:
CORRection:OPEN
e.g.
:
WrtCmd(“CORR:OPEN”)
:OPEN:STATe
is used to set the open correction function. Char ‘?’ is used to query the state of
the open correction.
command syntax
:
ON
CORRection:OPEN:STATe OFF
1
0
char ‘1’(49) :enable the open correction, which is equivalent to ON
char ‘0’(48): forbid the open correction, which is equivalent to OFF.
e.g.
:
WrtCmd(“CORR:OPEN:STAT ON”)
query syntax
:
CORRection:OPEN:STATe?