TH2826 Operation manual
Ver1.0
◇
74
The experiment instruments are shown in table 6-10
NO
The instrument name
requirement
100pF
1000pF
10000pF
10nF
0.1uF
1
Standard capacitance
1uF
0.02
%
(the loss D has been
known)
10
Ω
100
Ω
1k
Ω
10k
Ω
2
AC
standard resistor
100k
Ω
0.02
%
0.1
Ω
1
Ω
10
Ω
100
Ω
1k
Ω
10k
Ω
3
DC
standard resistor
100k
Ω
0.02
%
100
μ
H
1mH
10mH
4
Standard capacitance
100mH
0.02%
5
Frequency meter
(
0
~
1000
)
MHz
6
digital multimeter
0.5
%
7
Insulation
resistance meter
500V 10 levels
8
withstanding voltage and leakage meter
0.25kW
(
0
~
500
)
V
6.6.3Check the Function
The keys, the display and the terminals can work normally. The other functions are right.
6.6.4The test signal level
When the test signal level is tested, the digital multimeter is used. Adjust the range of the