![Tonghui Electronics TH2826/A Скачать руководство пользователя страница 71](http://html.mh-extra.com/html/tonghui-electronics/th2826-a/th2826-a_operation-manual_1126460071.webp)
TH2826 Operation manual
Ver1.0
◇
63
—20mARMS
+10
μ
ARMS
)
6.2.4Internal resistance mode
10/CC
(
Constant Current
)、
30
Ω
、
50
Ω
、
100
Ω
6.2.5Test signal level monitor
Figure 0-3
Mode
Range
Accuracy
Voltage
0V
RMS
—5V
RMS
±
(
10%×10mV
)
Current
0A
RMS
—100mA
RMS
±
(
10%×10
μ
A
)
6.2.4The maximum range of the test and display
Figure 0-4
Parameter
Displayed range
L
、
Lk
0.01nH
~
9.9999kH
C
0.0001pF
~
9.9999F
R
、
X
、
Z
、
DCR
0.1m
Ω
~
99.999M
Ω
Y
、
B
、
G
0.0001nS
~
99.999S
D
0.0001
~
9.9999
Q
0.0001
~
99999
θ
Deg -179.99°
~
179.99°
Rad -3.14159
~
3.14159
6.2.5The test voltage for DC resistance
±1.4VDC
Accuracy
:
±5%
Internal resistance
:
50
Ω
±5%
6.2.6 The internal DC voltage bias
-5v~+5V, ±(10%+10mV), 1mV by stepper
6.3Test accuracy
The relation of test accuracy and basic accuracy is shown as below
:
Test accuracy = Basic accuracy
×
C
×
D+E (1)
C
:
Speed factor
D
:
Cable length factor
E
:
Temperature factor
6.3.1 Basic accuracy
Basic accuracy is calculated by A, B in basic accuracy factor table
.
Note
:
(1) when frequency is larger than 1.001MHz
,
basic accuracy must multiply (f(MHz)+3)/4
;
Содержание TH2826/A