TH2826 Operation manual
Ver1.0
◇
112
return
:
<NR1><NL^END>
:SHORt
is used to perform the short correction data of 41 preset test points.
command syntax
:
CORRection:SHORt
e.g.
:
WrtCmd(“CORR:SHOR”)
:SHORt:STATe
is used to set the short correction function. Char ‘?’ is used to query the state
of the short correction.
Command syntax
:
ON
CORRection:SHORt:STATe OFF
1
0
char ‘1’(49):enable the short correction, which is equivalent to ON.
char ‘0’(48):forbid the short correction, which is equivalent to OFF.
e.g.
:
WrtCmd(“CORR:SHOR:STAT ON”)
query syntax
:
CORRection:SHORt:STATe?
return
:
<NR1><NL^END>
:LOAD:STATe
is used to set the load correction function. Char ‘?’ is used to query the state of
the load correction.
command syntax
:
ON
CORRection:LOAD:STATe OFF
1
0
char ‘1’(49):enable the load correction, which is equivalent to ON.
char ‘0’(48):forbid the load correction, which is equivalent to OFF.
e.g.
:
WrtCmd(“CORR
:
LOAD
:
STAT ON”)
query syntax
:
CORRection:LOAD:STATe?
return
:
<NR1><NL^END>
:LOAD:TYPE
is used to set the combination parameter function for the load correction. Char
‘?’is used to query the type of the combination parameter.
Fuction
:
CPD: set the function to Cp-D
LPRP: set the function to Lp-Rp
CPQ: set the function to Cp-Q
LSD : set the function to Ls-D
Содержание TH2826/A