TH2826 Operation manual
Ver1.0
◇
65
6.3.3Cable length factor
Cable length
0m
1m
(100kHz
max)1.5+0.015*f[kHz]
D
1
(100.1kHz
min)1.5+0.3*f[kHz]
6.3.4 Temperature factor
Temperature
Operation temperature=T
℃
E
0.1*basic accuracy*(T-20)
Basic accuracy table 1(0.010V to 0.049V)
When range>=100
Ω
Basic accuracy
|
[ ]
[ ] |
(
)
[ ]
B Zx
Range
A
Range
×
Ω −
Ω
= ± +
Ω
When range <100
Ω
Basic accuracy =
|
[ ] 10
[ ] |
(
)
[ ]
B Range
Zx
A
Range
×
Ω × −
Ω
= ± +
Ω
Zx:sample value of impedance Z
A,B:Basic accuracy factor
Accuracy factor of Z(%)
:
value A
,
B in group 1
Accuracy factor of
θ
(
°
)
:
value A
,
B in group 2
When
1.001
f
MHz
>=
,basic accuracy must multiply (f
[
MHz
]
+3)/4.
Table 6-5 Basic accuracy 1(0.010V to 0.049V)
*1
Range
20Hz-99.9Hz
100Hz-1kHz
1.001kHz-10kHz
10.01kHz-100kHz
100.1kHz-1MHz
1.001MHz-5MHz
1M
Ω
A=1.5 B=1
A=1 B=0.5
A=1 B=0.5
100k
Ω
A=2 B=0.3
A=0.5 B=0.1
A=0.5 B=0.1
A=0.8 B=0.5
A=0.5 B=0.2
A=0.5 B=0.2
A=1.5 B=0.5
10k
Ω
A=0.5 B=0.01
A=0.2 B=0.02
A=0.2 B=0.02
A=0.6 B=0.05
A=0.8 B=0.5
A=0.5 B=0.2
A=0.5 B=0.2
A=1.5 B=0.5
3k
Ω
A=0.5 B=0.01
A=0.2 B=0.02
A=0.2 B=0.02
A=0.6 B=0.05
Содержание TH2826/A