TH2826 Operation manual
Ver1.0
◇
99
char ‘0’( 48) is equivalent to OFF.
e.g.
:
WrtCmd(“FUNC:IMP:RANG:AUTO ON”); set the range to AUTO
query syntax
:
FUNCtion:IMPedance:RANGe:AUTO
?
return
:
<NR1><NL^END>
:Source MONitor:VAC
is used to set the monitor for voltage. Char ‘?’ can be used to query
the state of the voltage monitor.
Command syntax
:
ON (1)
FUNCtion:SMONitor:VAC OFF (0)
char ‘1’( 49) is equivalent to ON
char ‘0’( 48) is equivalent to OFF.
e.g.
:
WrtCmd(“FUNC:SMON:VAC ON”); set the voltage monitor ON.
query syntax
:
FUNCtion:SMONitor:VAC
?
return
:
<NR1><NL^END>
:Source MONitor:IAC
is used to set the current monitor. Char ‘?’can be used to query the
state of the current monitor.
command syntax
:
ON (1)
FUNCtion:SMONitor:IAC
OFF (0)
char ‘1’( 49) is equivalent to ON
char ‘0’( 48) is equivalent to OFF.
e.g.
:
WrtCmd(“FUNC:SMON:IAC ON”); set the current monitor ON
query syntax
:
FUNCtion:SMONitor:IAC
?
return
:
<NR1><NL^END>
: DEV<n>:MODE
is used to set the deviation measurement mode. Char ‘?’ is used to query
the state of the deviation measurement mode.
command syntax
: