TH2826 Operation manual
Ver1.0
◇
40
sweep
,
when user presses [TRIGGER] once
,
then the instrument sweeps curve.
PressTrigger
:
MAN
,
the zone displaysTrigger
:
EXT
,
which means external trigger
sweep
,
when the external interface or computer triggers once, then instrument
sweeps curve.
Press Trigger
:
EXT
,
the zone displaysTrigger
:
INT
,
which means internal trigger
sweep
,
after instrument sweeps curve, it performs next time automatically.
3)
Point
:
System sets the point as 101
,
201
,
401
,
801. The more points, the more precision of the
drawed figure, but the sweep time will be longer. The default point is 201.
Operation
:
press Point
:
to select the corresponding point.
4)
Curve setting
Press More 2/2 ,Coordinate
:
linearity
,
Ratio
:
auto
,
Load
,
Resweep and More 1/2 are
displayed.
When user selects sweep parameter as frequency and impedance
,
the instrument displays
some parameters of component
,
the detailed parameter is as below
:
Capacitance value when Ct
:
1kHz.
Capacitance value when Dt
:
1kHz.
Z
min
and f
s
,( point a in figure).
Z
max
and f
p
,( point A in figure).
51
.
2
×
−
≈
s
s
p
p
f
f
f
k
.
)
(
2
2
2
min
2
s
p
T
s
p
m
f
f
C
Z
f
f
Q
−
≈
π
.
Δ
F = fp – fs
As figure 3-15
: