TH2826 Operation manual
Ver1.0
◇
91
Command syntax
:
DISPlay:LINE”<string>”
<string> is the ASCII string(the maximum length is 20).
e.g.
:
WrtCmd(“DISP:LINE ”Resistor meas””);
query syntax
:
DISPlay:LINE?
return
:
<string><NL^END>
:ResultFONt
is used to set the font of the measurement result. Char’?’can be used to
query the current font of the measurement result.
Command syntax
:
DISPlay:RFONt <font>
The following value can be selected for <font>
:
LARGe:display the measurement result with the large size, the hold time is 12 ms.
TINY:display the measurement result with the small size, the hold time is 5 ms.
OFF:don’t display the measurement result, but the result can be read from the bus.
Query syntax
:
DISPlay:RFONt
?
return
:
<font><NL^END>
the following value can be selected for <font>
:
LARGE
TINY
OFF
8.1.2FREQuency subsystem command set
FREQuency
subsystem command set is used to set the test frequency. Char ‘?’can be used to
query the test frequency.
Command syntax
:
<value>
FREQuency MIN
MAX
<value>: NR1,NR2 or NR3 format with Hz,kHz or MHz as the
postfix
MIN: set the test frequency to 20Hz
MAX:set the test frequency to 200kHz
e.g.
:
WrtCmd(“FREQ 1KHZ”); set the test frequency to 1000Hz.
Query syntax
:
FREQuency
?
return
:
<NR3><NL^END>