TH2826 Operation manual
Ver1.0
◇
90
:RFONt
LARGe, TINY , OFF
:PAGE
is used to set the display page. Char ‘?’can can be used to query the current
page.
Command syntax
:DISPlay:PAGE <page name>
The following value can be selected for <page name>:
MEASurement: set the page to LCR Measurement display.
BNUMber: set the page to BIN Number display.
BCOUnt: set the page to BIN Count.
LIST: set the page to List Sweep.
MSETup: set the page to Measurement setup.
CSETup: set the page to Correction.
LTABle: set the page to Limit Table
LSETup: set the page to List Setup
SYSTem: set the page to System Setup
FLISt: set the page to File List.
E.g.:WrtCmd(“ISP:PAGE MEAS”); set the page to LCR Measurement display.
Query syntax
:
DISPlay:PAGE?
return
:
<page name><NL^END>
The following value can be selected for <page name>:
<LCR MEAS DISP>: the current page is LCR Measurement display
<BIN No. DISP>: the current page is BIN Number display
<BIN COUNT DISP>: the current page is BIN Count
<LIST SWEEP DISP>:
the current page is List Sweep
<MEAS SETUP>:
the current page is Measurement setup
<CORRECTION>:
the current page is Correction
<LIMIT TABLE SETUP>: the current page is Limit Table
<LIST SWEEP SETUP>: the current page is List Setup
<SYSTEM SETUP>:
the current page is System Setup
<FILE LIST>:
the current page is File List
:LINE
is used to set the measurement theme. The maximum length of the theme string
is 20. Char ‘?’can can be used to query the current measurement theme.The theme
string will be displayed in the file list page, and saved as the file name.