TH2826 Operation manual
Ver1.0
◇
103
return
:
<NR3>[,<NR3> *]<NL^END>
note:the DC bias voltage must be 0 V, 1.5 V and 2 V, or the return value will be “error”..
:BIAS:CURRent
is used to clear the DC bias current in the list sweep points before and set the
DC bias current.
Command syntax
:
LIST:BIAS:CURRent <value>[,<value> *]
note
:
” *” shows that the sweep points are 10 points at most.
<value>:NR1
,
NR2 or NR3 format
e.g.
:
WrtCmd(“LIST:BIAS:CURR 100MA”) ;set the DC current of the sweep point 1
to100mA
WrtCmd(“LIST:BIAS:CURR 1E-2,2E-2,3E-2,4E-2”) ;set the DC current of the sweep
point 1,2,3 and 4 to 10mA
,
20mA
,
30mA and 40mA
quety syntax
:
LIST:BIAS:CURRent?
return
:
<NR3>[,<NR3> *]<NL^END>
note: because TH2826 and TH2826 have no internal DC bias current source, the instrument
must be used with the external DC bias source(TH1773, must be bought.) together.
:MODE
is used to set the list sweep mode.
command syntax
:
SEQuence
LIST:MODE
STEPped
SEQuenc: sequence mode
STEPped: stepped mode
e.g.
:
WrtCmd(“LIST:MODE SEQ”)
query syntax
:
LIST
:
MODE
?
return
:
SEQ
<NL^END>
STEP
:BAND<n>
is used to set the limit data in the list sweep table.
command syntax
:
LIST:BAND<n><parameter>[,<low limit n>,<high limit n>]
<n> 1 to 10
(
NR1 format
):
sweep point in the line n
<parameter>: A compare the main parameter of the measurement result with the
high limit and the low limit.
B compare the sencondary parameter of the measurement result