TH2826 Operation manual
Ver1.0
◇
55
of the test error is electromagnetic coupling between the test leads. TH2826 employs a
four-terminal pair measurement configuration to eliminate the effect of the coupling. Because
the current in HD, LD and the one in the shielding terminal are the same size, but the opposite
direction, the magnetic fields are eliminated.
5.5Inductance test example
measurement conditions:
function: Ls-Q
frequency:5.5kHz
level:1.5Vrms
internal resistance:100
Ω
Procedures:
1) Turn on the TH2826.
2) Set the basic parameters:
a)
Press [LCRZ], LCR MEAS DISP page will be displayed.
b)
Move the cursor to FUNC field by using the direction keys on the front panel.
The FUNC field displays Cp-D. At this time, Cp-D, Cp-Q, Cp-G, Cp-Rp and
MORE 1/6 will be displayed in the softkey area, which is on the right of the
screen.
c)
Press MORE 1/6. Cs-D, Cs-Q, Cs-Rs and MORE 2/6 will be displayed.
d)
Press MORE 2/6. Lp-D, Lp-Q, Lp-G, Lp-Rp and MORE 3/6 will be displayed.
e)
Press MORE 3/6.Ls-D,Ls-Q,Ls-Rs and MORE 2/6 will be displayed.
f)
Press Ls-Q to select Ls-Q test function.
g)
Move the cursor to FREQ. the current frequency displayed is 1.0000kHz.
h)
Press key[5][.][5], +5.5 will be displayed in the bottom line of the screen. The
softkey area will display the usable units (Hz, kHz and MHz).Press kHz, the
FREQ field will be changed to 5.5000kHz.
i)
Move the cursor to LEVEL field. The current level displayed is 1.000V.