TH2826 Operation manual
Ver1.0
◇
13
Operation step of range setting
:
1)
Move cursor to Range. The following softkeys will be displayed.
AUTO The softkey is used to set range AUTO mode
HOLD The softkey is used to switch AUTO to HOLD. When the range is set as
HOLD mode
,
the range is locked in the current range. And the current
range is displayed on the Range zone.
↑
(+)
The softkey is used to increase range in the mode (HOLD).
↓
(-) The softkey is used to decrease range in the mode (HOLD)
。
2)
Use softkey to set range.
3.1.3Frequency
The test frequency of TH2826 is from 20 Hz to 5MHz, and test frequency of TH2826 A is
from 20 Hz to 2MHz, the resolution is
:
0.01 Hz. When the test function is set as DCR
,
Frequency zone displays“----”
Operation step of frequency setting
:
TH2826 has two kinds of frequency setting. One is using softkey
,
the other is using numeric
key.
1)
Move cursor to Frequency zone. The following softkeys will be displayed.
↑
(++)
The key is used to increase frequency roughly. Press this key, the frequency will be
increased to the next 10 times frequency point after 20 Hz.
↑
(+)
The following softkeys will be displayed exactly. Press this key, the frequency will
be increased to a higher frequency point. 10 frequency points can be set among 10
times frequency point.
↓
(-)
The key is used to decrease frequency exactly. Press this key, the frequency will be
increased to a lower frequency point. 10 frequency points can be set among 10 times
frequency point. The frequency point can be set is the same as
↑
(+)
。
↓
(--)
The key is used to decrease frequency roughly. Press this key, the frequency will be
increased to the next 10 times frequency point. The frequency point can be set is the
same as
↑
(++).
2)
Use softkey or numberic key to select or set frequency. When using numeric key to input
the frequency value, the frequency units are (Hz, kHz, and MHz), which can be used to
input unit and data. When using [ENTER] to input frequency
,
the default unit is Hz.
3.1.4Test level
The test level of TH2826 is set as the valid value of sine wave. The frequency of sine signal is
Содержание TH2826/A