6.
Load the plate into the instrument, then tap
Start
.
CAUTION!
The instrument should be used by trained operators who have been
warned of the moving parts hazard.
Calibration complete
is displayed when the run is complete.
7.
Tap
View results
4
Details
.
8.
Review the plot.
Calibration results that pass show uniform signals with peaks that are aligned
with the dye wavelength.
Peak filter
Filter wavelength (nm)
[1]
Excitation
Emission
x1-m1
470 ± 15
520 ± 15
x2-m2
520 ± 10
558 ± 11
x3-m3
550 ± 11
586 ± 10
x4-m4
580 ± 10
623 ± 14
x5-m5
640 ± 10
682 ± 14
x6-m6
662 ± 10
711 ± 12
[1]
The central wavelengths are the optimized wavelengths.
Figure 14 An example of a dye calibration plot
The peaks for your dye may align with a different filter set.
Chapter 7
Calibrate and verify instrument performance
Calibrate custom dyes
7
QuantStudio
™
6 Pro and 7 Pro Real-Time PCR Systems User Guide
103