4
Done
to remove the Auto Delta setting.
Note:
The Auto Delta setting is removed from the
Method
tab.
1
2
1
Indicates that this is the
Method
tab where Auto Delta settings can be removed.
2
Indicates that Auto Delta is assigned. Tap to remove the Auto Delta settings.
Configure ramp rates
The ramp rate is how quickly the temperature changes from one stage or step to
another stage or step. The maximum ramp rate for the instrument is 6.5°C/second.
1.
In the
Method
tab, tap
Edit
4
Advanced
4
Ramp rates
.
Note:
Use or to navigate through the method if it is not all displayed.
2.
Tap a ramp rate, then edit the value.
3.
Tap
Done
.
Add or remove a pause
A pause allows you to open the drawer to add additional reagents to the plate.
1.
In the
Method
tab, tap
Edit
4
Advanced
4
Add pause
.
Note:
Use or to navigate through the method if it is not all displayed.
2.
Tap
(Pause)
.
3.
Enter a temperature and the cycle.
A pause in the thermal cycling protocol is denoted with a . The cycle of the
pause and the temperature of the pause are also denoted.
4.
Tap
(Pause)
to remove the pause.
5.
Tap
Done
.
Chapter 4
Edit a plate before starting a run
Options to edit the method of a plate file
4
QuantStudio
™
6 Pro and 7 Pro Real-Time PCR Systems User Guide
59