(continued)
Peak
filter
Color
Filter wavelength (nm)
[1]
System dyes
Example
custom dyes
Excitation
Emission
x5-m5
Red
640 ± 10
682 ± 14
MUSTANG
PURPLE
™
dye,
Cy
®
5
LIZ
™
dye
x6-m6
Deep-
Red
662 ± 10
711 ± 12
None
[3]
Cy
®
5.5 dye
[1]
The central wavelengths are the optimized wavelengths.
[2]
The HEX
™
dye and TET
™
dye from Thermo Fisher Scientific fall within the emission wavelength range of the
system, therefore they can be added and adapted for use on the instrument.
[3]
This filter set currently does not support any dyes supplied by Thermo Fisher Scientific.
Filters
Wavelength
(nm)
1
2
3
4
5
x1-m1 x2-m2 x3-m3 x4-m4 x5-m5 x6-m6
Emission
Spectra
1
x1-m1—FAM
™
dye, SYBR GREEN
™
dye
2
x2-m2—VIC
™
dye
3
x3-m3—ABY
™
dye, NED
™
dye, Cy
®
3,
TAMRA
™
dye
4
x4-m4—JUN
™
dye, ROX
™
dye, Texas Red
™
dye
5
x5-m5—Cy
®
5, MUSTANG PURPLE
™
dye
Custom dyes
The instrument can run assays designed with custom dyes. Custom dyes include the
following options:
• Dyes that are not supplied by Thermo Fisher Scientific.
• Dyes or formulations of dyes that are not system dyes for the instrument.
To use a custom dye on the instrument, review the following requirements.
• Calibrate the instrument for the custom dye (see “Calibrate custom dyes” on
• Ensure that the custom dye excites between 455–672 nm and emits between
505–723 nm.
• Select a custom dye that does not overlap with other dyes used in the run (see
• The custom dye for use in calibration must be attached to the 5' end of a short
DNA oligonucleotide consisting of the first two bases of the probe sequence
without a quencher at the 3' end.
Chapter 1
Product information
Instrument hardware description
1
16
QuantStudio
™
6 Pro and 7 Pro Real-Time PCR Systems User Guide