5.
When the run is complete and the screen displays
Complete
, tap
View results
to confirm the status of the run.
Verification status
Action
Passed
Instrument is ready for use.
Failed
See “Troubleshoot verification failure” on page 96.
6.
In the
RNase P Verification Status
screen, tap one of the following options.
Option
Description
Accept results
Save the results to the instrument.
Reject results
Deletes the RNase P verification results.
Export results
Exports the results.
7.
Unload the plate from the instrument.
CAUTION!
PHYSICAL INJURY HAZARD.
During instrument operation, the
plate temperature can reach 100°C. Allow it to cool to room temperature before
handling.
Note:
Use Smart Help if the instrument does not eject the plate (see “Request
technical support with Smart Help” on page 159).
The status of the RNase P verification is displayed on the
Status and Reminders
screen (see “View the calibration status on the instrument” on page 81).
IMPORTANT!
RNase P plates can only be used one time. Do not reuse the RNase P
plate.
Chapter 7
Calibrate and verify instrument performance
Perform verification using RNase P plates
7
QuantStudio
™
6 Pro and 7 Pro Real-Time PCR Systems User Guide
95