Perform ROI/uniformity, background, and dye calibrations
Calibration workflow
Perform an ROI/uniformity calibration
You are automatically prompted to perform background calibration.
▼
Perform a background calibration
Perform any time that ROI/uniformity calibrations are current.
▼
Perform system dye calibrations
Perform any time that ROI/uniformity and background calibrations are current.
▼
(
Optional
) Perform custom dye calibrations
(see “Calibrate custom dyes” on page 98)
Perform any time that ROI/uniformity and background calibrations are current.
Prepare a calibration plate
Required materials to prepare calibration plates
Note:
We recommend calibrating with all Spectral Dye Calibrations Plates available
for your block configuration even if you are not using all the dyes in the plates. This is
recommended for optimal system performance.
Unless otherwise indicated, all materials are available through
.
"MLS" indicates that the material is available from
major laboratory supplier.
Item
Source
Calibration plates
ROI/Uniformity plate
Background calibration plate
Dye calibration plates
Equipment and consumables
Centrifuge with plate adapter; buckets cleaned before use
MLS
Powder
‑
free gloves
MLS
Safety glasses
MLS
Chapter 7
Calibrate and verify instrument performance
Perform ROI/uniformity, background, and dye calibrations
7
84
QuantStudio
™
6 Pro and 7 Pro Real-Time PCR Systems User Guide