Maintain the instrument
■
Install a block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
■
Change the block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
■
Remove the block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
■
Decontaminate the sample block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
■
Replace the instrument fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
■
Power on or off, store, and move . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
IMPORTANT!
This chapter contains user maintenance procedures for the
instrument. Procedures other than those described in this document must be
performed by a qualified Thermo Fisher Scientific service representative.
Install a block
This section describes installing a block when there is no block in the instrument. To
change a block, see “Change the block” on page 144.
A block can be installed by any user. This function is not restricted to administrators.
Each combination of a block and an instrument has its own calibration status. To
determine the calibration status of a block on an instrument, see “View the calibration
status and set reminders” on page 81. To calibrate a block on a specific instrument,
see Chapter 7, “Calibrate and verify instrument performance”.
1.
Tap
(Settings)
4
Change block
.
2.
Insert the block and heated cover into the instrument drawer.
The instructions are displayed on the screen.
3.
Tap
Install
.
The instrument displays one of the following options, depending on the calibration
status:
• Confirmation that the block was installed successfully and the expiration date of
the current calibration.
• A warning that the block is not calibrated on the instrument.
• A warning that the calibration of the block on the instrument is expired.