• Calibration name
• Result
• Date
• Whether this calibration is current
• The operator
• Reagent name, part number, lot
number, and expiration date
• Comments.
4.
Tap
Detail
.
Calibration
Example results indicating successful calibration
ROI
Note:
Select the desired
filter combination from the
Filter Set dropdown list.
Green circles around all wells and bright well centers.
Uniformity
Signals from each well following a uniform trend.
Background
Few, if any, signals with abnormally high fluorescence.
Dye
Signals from each well following a uniform trend, and
each dye peaks at the correct filter.
5.
(Optional)
For Uniformity, Background, and Dye calibration images, tap on an
individual well on the right side to see the results for a single well.
Chapter 7
Calibrate and verify instrument performance
Perform ROI/uniformity, background, and dye calibrations
7
88
QuantStudio
™
6 Pro and 7 Pro Real-Time PCR Systems User Guide