2.
Tap
(Data collection location)
to designate a step as a data collection
location or to remove data collection from the step.
The
(Data collection location)
button functions as an on-off switch.
2
1
1
Data collection is off
2
Data collection is on
3.
Tap
Done
.
Set up or remove VeriFlex
™
Zones
VeriFlex
™
Zones are only available for 96-well blocks.
• The QuantStudio
™
6 Pro System has three VeriFlex
™
Zones.
• The QuantStudio
™
7 Pro System has six VeriFlex
™
Zones.
Note:
For more information about compatibility of plate files with VeriFlex
™
Zones
between the two instruments, see “VeriFlex
1.
In the
Method
tab, tap
Edit
4
Advanced
4
VeriFlex
™
Zones / Auto Delta
.
Note:
Use or to navigate through the method if it is not all displayed.
Note:
The PCR stage provides the option of VeriFlex
™
Zones or the Auto Delta
feature. Ensure that the
VeriFlex
™
Zones
option is selected.
1
Figure 8 VeriFlex
™
Zones and Auto Delta settings
1
VeriFlex
™
Zones
is selected
Chapter 4
Edit a plate before starting a run
Options to edit the method of a plate file
4
QuantStudio
™
6 Pro and 7 Pro Real-Time PCR Systems User Guide
57