Orion Star Series Meter
ISE Technique
ISE Calibration for 1-5 Points
1 . Prepare the electrode and standard for use according to the electrode user
guide or instruction manual .
2 . Press
until the arrow icon is pointing to the
ISE
measurement line .
3 . Press
.
4 . Rinse the electrode and place into the standard .
5 . Wait for
ISE
to stop flashing . Press
/
and
to change the value of
the standard .
6 . Press
to proceed to the next calibration point and repeat steps 4-6 or press
to save the calibration .
7 . The slope will be displayed before the meter returns to the measurement mode .
is displayed in the lower field and the actual electrode slope, in
mV
, is
displayed in the main field .
Note:
For a 2 or more point calibration the meter will automatically proceed
to measurement mode after 3 seconds. For a 1 point calibration the meter will
allow you to edit the slope then press
to proceed to measurement mode.
s
Note:
If editing a negative number for slope:
s
a.
Press
until no digit is blinking and the arrow icon is blinking.
b.
Press
/
to change the sign of the slope .
c.
Press
to save.