![Thermo Orion Star Скачать руководство пользователя страница 100](http://html.mh-extra.com/html/thermo/orion-star/orion-star_user-manual_1104232100.webp)
Thermo Electron Corporation
XIII-18
Orion Star Series Meter
Care and Maintenance
Example of Conductivity Calibration
Cal. Pt. 1
Point 1
Conductivity 1
Conductivity
100.0 µS/cm
Conductance 1
Conductance
10.0 mS
Temp. 1
Temperature
25.0 C
Auto/Manual
Cal. Type:
Auto
Blank line
Cal. Pt. 2
Point 2
Conductivity 2
Conductivity
1413 uS/cm
Conductance 2
Conductance
32.2 mS
Temp. 2
Temperature
25.0 C
Auto/Manual
Cal. Type:
Auto
Blank line
Cal. Pt. 3
Point 3
Conductivity 3
Conductivity
12.9 mS/cm
Conductance 3
Conductance
67.2 mS
Temp. 3
Temperature
25.0 C
Auto/Manual
Cal. Type: Auto
Blank line
Cal. Pt. 4
Point 4
Conductivity 4
Conductivity
111.9 mS/cm
Conductance 4
Conductance
100.0 mS
Temp. 4
Temperature
25.0 C
Auto/Manual
Cal. Type: Aut
Blank line
Cal. Pt. 5
Point 5
Conductivity 5
Conductivity
200.0 mS/cm
Conductance 5
Conductance
150.5 mS
Содержание Orion Star
Страница 1: ...Orion Star Series Meter User s Guide Analyze Detect Measure Control ...
Страница 12: ...Thermo Electron Corporation III 4 Orion Star Series Meter NOTES ...
Страница 56: ...Thermo Electron Corporation VIII 14 Orion Star Series Meter Dissolved Oxygen Technique NOTES ...
Страница 108: ...Thermo Electron Corporation XIV 6 Orion Star Series Meter NOTES ...