Thermo Electron Corporation
VIII-10
Orion Star Series Meter
Dissolved Oxygen Technique
2 . To initiate the Winkler titration calibration, press
. The
“cal” indicator will light up .
3 . While the electrode signal is stabilizing, the meter will
flash a dissolved oxygen reading (based upon the current
calibration data) in the main field . Once the input has
stabilized, the concentration value, which corresponds to
saturation at the current temperature, pressure and salinity,
will be displayed .
4 . The concentration value should then be adjusted to the
value that was obtained via the Winkler titration . Use
/
and
to enter value . Press
to accept the
concentration value .
5 . The slope value (
) will then be displayed for three
seconds . The meter will then return to measure mode .
Probe Zero Calibration
If the expected sample range is at the lower end of the
dissolved oxygen scale, you may want to perform a zeroing
of your probe . If the probe exhibits any residual current, this
calibration will compensate for it . A Probe Zero calibration
should always follow as the second point determination of a
two-point calibration . To remove the effects of a Probe Zero
calibration, perform an Air, Water or Winkler calibration as a one-
point calibration . To perform a Probe Zero calibration, follow the
steps outlined below .
Содержание Orion Star
Страница 1: ...Orion Star Series Meter User s Guide Analyze Detect Measure Control ...
Страница 12: ...Thermo Electron Corporation III 4 Orion Star Series Meter NOTES ...
Страница 56: ...Thermo Electron Corporation VIII 14 Orion Star Series Meter Dissolved Oxygen Technique NOTES ...
Страница 108: ...Thermo Electron Corporation XIV 6 Orion Star Series Meter NOTES ...