VIII-13
Orion Star Series Meter
Thermo Electron Corporation
Dissolved Oxygen Technique
When using the AutoStir™ BOD Probe, pressing the button on the probe will
start the AUTO-READ measurement .
c . If you are in Timed Measurement mode then the meter will start making
measurements as soon as it goes into measurement mode at the frequency
selected in setup . It will automatically Log & Print each measurement . If
you are using the benchtop meter and the stirrer control is enabled then
pressing
will start the stirrer . Pressing
again will turn off the stirrer .
3 . Remove the electrode from the sample and rinse with deionized water, then
place it in your next sample and repeat step 2 .
4 . When all samples have been measured, rinse the electrode with deionized
water and blot dry . Consult your electrode manual for proper electrode storage
instructions .