Thermo Electron Corporation
IX-8
Orion Star Series Meter
The will flash until the reading is stable . Once the reading is stable you
can Log & Print the measurement by pressing the
. If you are recording
the data into a notebook, record the pH as well as the temperature at which
the pH reading is taken . If the stirrer is in motion, press
to turn off the
stirrer before removing the stirrer from the sample .
b . If you are in AUTO-READ™ mode then press
to start the measurement .
Once the reading is stable the meter will automatically Log & Print the
results and freeze the display . If the stirrer is enabled, then the stirrer will
turn on when
is pressed and turn “off” once the reading has stabilized .
When using the AutoStir™ BOD Probe, pressing the button on the probe will
start the AUTO-READ measurement .
c . If you are in Timed Measurement mode then the meter will start making
measurements as soon as it goes into measurement mode at the frequency
selected in setup . It will automatically Log & Print each measurement . If
you are using the benchtop meter and the stirrer control is enabled then
pressing
will start the stirrer . Pressing
again will turn off the stirrer .
3 . Remove the electrode from the sample and rinse with deionized water, then
place it in your next sample and repeat step 2 .
4 . When all samples have been measured, rinse the electrode with deionized
water and blot dry . Consult your electrode manual for proper electrode storage
instructions .
Содержание Orion Star
Страница 1: ...Orion Star Series Meter User s Guide Analyze Detect Measure Control ...
Страница 12: ...Thermo Electron Corporation III 4 Orion Star Series Meter NOTES ...
Страница 56: ...Thermo Electron Corporation VIII 14 Orion Star Series Meter Dissolved Oxygen Technique NOTES ...
Страница 108: ...Thermo Electron Corporation XIV 6 Orion Star Series Meter NOTES ...