Thermo Electron Corporation
X-2
Orion Star Series Meter
ISE Technique
ISE Setup Options
Description
Display
Range of Values
Default
Value
Method
Specific
ISE Resolution
ISE
rES
1
1, 2, 3
ISE Resolution in
Significant Digits
1
Yes
ISE Units
ISE
UnIt
PPb
M, mG/L, per, PPb, nOnE
ISE Units Displayed
PPb
Yes
ISE Calibration Standard
Concentration Range
ISE
rAng
HIgH
LOw, HIgH
Stability criteria used
during ISE Calibration
HIgH
Yes
ISE Auto-Blank Correction
ISE
nLIn
AUtO
AUtO, OFF
Enables or Disables
Calibration Auto-Blank
AUtO
Yes
1 . In the measurement mode press
.
2 . Press
/
to scroll through the setup menu until
is displayed on the
top line .
3 . Press
to accept the selection and move to the middle line .
4 . Press
/
to scroll through the following options:
•
= Resolution
•
= ISE Measurement units
•
M
= Molar
•
mG/L
= mg/L
•
PEr
= Percent
•
PPb
= Parts per Billion
•
nOnE
= No units