V-13
Orion Star Series Meter
Thermo Electron Corporation
Menu Setup
Calibration Alarm Setup
The following table describes in greater detail the abbreviations shown on the
display screen .
Description
Display
Range of Values
Default
Value
Method
Specific
pH Calibration Alarm
Setting
dUE
PH
0
0000 - 9999
Time in Hours; 0000 = Cal
Alarm Off
0000
Yes
ORP Calibration Alarm
Setting
dUE
OrP
0
0000 - 9999
Time in Hours; 0000 = Cal
Alarm Off
0000
Yes
ISE Calibration Alarm
Setting
dUE
ISE
0
0000 - 9999
Time in Hours; 0000 = Cal
Alarm Off
0000
Yes
Conductivity Calibration
Alarm Setting
dUE
COnd
0
0000 - 9999
Time in Hours; 0000 = Cal
Alarm Off
0000
Yes
Dissolved Oxygen
Calibration Alarm Setting
dUE
dO
0
0000 - 9999
Time in Hours; 0000 = Cal
Alarm Off
0000
Yes
The following steps enable calibration alarms:
1 . In the measurement mode, press
.
2 . Press
/
to scroll through the setup menu until
is displayed on the top line .
3 . Press
to accept the selection and move the arrow icon
to the middle line .
4 . Press
/
to scroll through the measurement modes .
Содержание Orion Star
Страница 1: ...Orion Star Series Meter User s Guide Analyze Detect Measure Control ...
Страница 12: ...Thermo Electron Corporation III 4 Orion Star Series Meter NOTES ...
Страница 56: ...Thermo Electron Corporation VIII 14 Orion Star Series Meter Dissolved Oxygen Technique NOTES ...
Страница 108: ...Thermo Electron Corporation XIV 6 Orion Star Series Meter NOTES ...