Thermo Electron Corporation
VIII-4
Orion Star Series Meter
Dissolved Oxygen Technique
• The user may disable the automated process by selecting the
setting
rather than
.
• DO Manual Salinity Correction Factor Setting adjusts the dissolved oxygen
concentration in mg/L based on the samples salinity . Use the following table
to select an appropriate salinity factor . A typical value for Sea Water is 35,
but does vary considerably based on location .
Conductivity
in mS/cm
Salinity
Value
Conductivity
in mS/cm
Salinity
Value
Conductivity
in mS/cm
Salinity
Value
5
3
20
13
35
25
6
4
21
14
36
25
7
4
22
15
37
26
8
5
23
15
38
27
9
6
24
16
39
28
10
6
25
17
40
29
11
7
26
18
42
30
12
8
27
18
44
32
13
8
28
19
46
33
14
9
29
20
48
35
15
10
30
21
50
37
16
10
31
22
52
38
17
11
32
22
54
40
18
12
33
23
56
42
19
13
34
24
Salinity depending on the conductivity at 20 ˚C
ppt = Parts per Thousand Salinity
This table was calculated from the International Oceanographic Tables . International Oceanographic
Tables, Vol . 1, National Institute of Oceanography of Great Britian, Womley, Godaming, Surrey, England
and Unesco, Paris 1971
Содержание Orion Star
Страница 1: ...Orion Star Series Meter User s Guide Analyze Detect Measure Control ...
Страница 12: ...Thermo Electron Corporation III 4 Orion Star Series Meter NOTES ...
Страница 56: ...Thermo Electron Corporation VIII 14 Orion Star Series Meter Dissolved Oxygen Technique NOTES ...
Страница 108: ...Thermo Electron Corporation XIV 6 Orion Star Series Meter NOTES ...