Thermo Electron Corporation
XI-4
Orion Star Series Meter
mV and ORP Technique
2 . Place the electrode in your sample .
a . If you are in continuous measurement mode, the instrument will start
measuring immediately . If you are using the benchtop meter & the stirrer
control is disabled then pressing
will start the stirrer . The
Rel mV
will
flash until the ready is stable . Once the ready is stable you can Log & Print
the measurement by pressing
. If the stirrer is on press
to turn off
the stirrer before removing the stirrer from the sample .
b . If you are in AUTO-READ™ mode then press
to start the measurement .
Once the ready is stable the meter will automatically Log & Print the results
and freeze the display . If the stirrer is enabled then the stirrer will turn on
when
is pressed and turn off once the reading has stabilized .
c . If you are in timed measurement mode then the meter will start making
measurements as soon as it goes into measurement mode at the frequency
set in setup . It will automatically Log & Print each measurement . If you are
using the benchtop meter & the stirrer control is enabled then pressing
will start the stirrer . Pressing
again will turn off the stirrer .
3 . Remove the electrode from the sample and rinse with deionized water, then
place it in your next sample and repeat step 2 .
4 . Once you have finished measuring samples, rinse the electrode with deionized
water and blot dry . Then consult your electrode manual for proper electrode
storage instructions .
Содержание Orion Star
Страница 1: ...Orion Star Series Meter User s Guide Analyze Detect Measure Control ...
Страница 12: ...Thermo Electron Corporation III 4 Orion Star Series Meter NOTES ...
Страница 56: ...Thermo Electron Corporation VIII 14 Orion Star Series Meter Dissolved Oxygen Technique NOTES ...
Страница 108: ...Thermo Electron Corporation XIV 6 Orion Star Series Meter NOTES ...