![Thermo Orion Star Скачать руководство пользователя страница 41](http://html.mh-extra.com/html/thermo/orion-star/orion-star_user-manual_1104232041.webp)
VII-5
Orion Star Series Meter
Thermo Electron Corporation
pH Technique
pH Measurements
1 . Rinse the electrode, as well as ATC probe and stirrer if being used, in deionized
water . Blot dry with a lint-free tissue . Do not wipe glass electrodes as static
charge will build up on the sensor .
2 . Place the electrode in your sample .
a . If you are in continuous measurement mode, the instrument will start
measuring immediately . If you are using the benchtop meter & the stirrer
control is enabled then pressing
will start the stirrer .
The will flash until the reading is stable . Once the reading is stable you
can Log & Print the measurement by pressing the
. If you are recording
the data into a notebook, record the pH as well as the temperature at which
the pH reading is taken . If the stirrer is in motion, press
to turn off the
stirrer before removing the stirrer from the sample .
b . If you are in AUTO-READ™ mode then press
to start the measurement .
Once the reading is stable the meter will automatically Log & Print the
results and freeze the display . If the stirrer is enabled, then the stirrer will
turn on when
is pressed and turn “off” once the reading has stabilized .
When using the AutoStir™ BOD Probe, pressing the button on the probe will
start the AUTO-READ measurement .
c . If you are in Timed Measurement mode then the meter will start making
measurements as soon as it goes into measurement mode at the frequency
selected in setup . It will automatically Log & Print each measurement . If
you are using the benchtop meter and the stirrer control is enabled then
pressing
will start the stirrer . Pressing
again will turn off the stirrer .
Содержание Orion Star
Страница 1: ...Orion Star Series Meter User s Guide Analyze Detect Measure Control ...
Страница 12: ...Thermo Electron Corporation III 4 Orion Star Series Meter NOTES ...
Страница 56: ...Thermo Electron Corporation VIII 14 Orion Star Series Meter Dissolved Oxygen Technique NOTES ...
Страница 108: ...Thermo Electron Corporation XIV 6 Orion Star Series Meter NOTES ...