IX-5
Orion Star Series Meter
Thermo Electron Corporation
Conductivity Technique
4 . Press
/
to scroll through the following Conductivity settings:
•
= Temperature Compensation Selection . The users may elect to turn
temperature compensation
, select linear
temperature
compensation, or a non-linear
temperature compensation for
natural and ultra-pure water .
•
= Temperature coefficient for
(linear) temp compensation
expressed in % /C .
•
= TDS Factor for Total Dissolved Solids measurement
•
= Conductivity Cell Constant (nominal cell constant used for
Auto-Calibration)
•
= Reference temperature for temperature compensation . Options
are degrees C,
degrees C, or
degrees C .
•
= Conductivity cell type (Planar or Conventional)
5 . Press
to accept the selection and move the arrow icon to the bottom line .
6 . Press
/
and
to edit value .
7 . Press
to accept the selection and move the arrow to the top line .
8 . Repeat steps 3-7 to edit conductivity settings as needed .
9 . Press
to return to measurement mode .