Thermo Electron Corporation
XIII-2
Orion Star Series Meter
Care and Maintenance
Error
Code
Description
Press
to Clear
Cause
Trouble Shooting
E034
Input Failure
Yes
Hardware problem or
BNC shorting cap not
attached during self test
Attach BNC shorting cap
and return the self test
E035
Password Wrong
Yes
User entered
incorrect password
Re-enter the password
If you have forgotten your
password contact Customer
Service
E036
User Entered
Wrong Value
Yes
User entered value is
out of range
Re-enter value. Check
specifications for range of
correct values
E038
Log Full
Yes
User setup as delete on
download & data log
is full
Download the datalog or
change setup to delete on
download off. This will
start over writing the oldest
log values.
E105
Auto-Buffer Failure
Yes
Auto-buffer not
recognized, will
only happen in remote
control
Contact Customer Service
E??
Memory Error
Yes
A memory read, write,
verify, or checksum error
This is ANY error message
starting with E. The 2
nd
2 digits refer to what
memory section failed.
Clear error and if repeats
return for service.
F01
No DO board in DO
meter
Yes
Meter is missing DO board Return for service
F02
No FET board in
FET meter
Yes
Meter is missing FET
board
Return for Service
F05
An option board is
installed and none
is required
Yes
Meter has a DO or FET
board installed and it’s not
used in this model
Return for service
Содержание Orion Star
Страница 1: ...Orion Star Series Meter User s Guide Analyze Detect Measure Control ...
Страница 12: ...Thermo Electron Corporation III 4 Orion Star Series Meter NOTES ...
Страница 56: ...Thermo Electron Corporation VIII 14 Orion Star Series Meter Dissolved Oxygen Technique NOTES ...
Страница 108: ...Thermo Electron Corporation XIV 6 Orion Star Series Meter NOTES ...