XIII-9
Orion Star Series Meter
Thermo Electron Corporation
Care and Maintenance
Example printouts
The examples below show the printer version of the RS232 output . If the user
has selected the computer output, then the data will be presented in the same
order, but only the raw data (no labels or units) will be given in a comma separated
format . User fields will not be output in the computer format . A user field is a label
with a line after it, such as “Sample # ____________” . In the example printouts, this
computer output text will be in red .
Example of a pH printout using “printer” setting:
3-Star BENCHTOP
pH
Meter S/n
123456
SW rev
1.29
Method #
5
06-29-2004
09:05:27
pH 10.45 pH
mV -183.3 mV
Temperature
25.0 C
Calibration
# 3
Log # 2
Sample # ____________
Example of a pH printout using “computer” setting:
112,123456,28,5,06-29-2004 09:05:27,-183.3,mV,10.45,pH,25.0,C,3,2
Note that the sample # and operator lines are not shown in the computer printout .
Blank lines are also skipped . Units are given directly after a measurement . If a
measurement does not have units, then this field will be blank .