Thermo Electron Corporation
X-1
Orion Star Series Meter
Chapter X
ISE Technique
Definition of Terms and Concepts
• ISE Resolution selects the number of significant digits displayed in
measurement mode . In addition, the resolution selection is used by SMART
STABILITY™ & SMART AVERAGING™ algorithms to optimize the accuracy,
precision, and response of the ISE measurement .
• ISE Units denote the units to be displayed in measurement & calibration . Both
measurement and calibration must be performed in the same units .
• ISE Calibration Standard Concentration Range selects the stability criteria to
be used in the ISE calibration . For most calibrations the
setting is the
best option . The
setting can improve the accuracy of low concentration
measurements by allowing the sensor a longer amount of time to stabilize in the
calibration standard .
• ISE Auto-Blank Correction uses Thermo’s patented Auto-Blank algorithm to
compensate for sensor non-linearities in low-level measurements . Since these
require the use of a set of non-linear equations that can only be calculated
numerically, the user cannot analytically verify the calibration . In applications
were analytical verification is a requirement, the user can choose the turn the
Auto-blank correction
.
Содержание Orion Star