ChromQuest Diagnostics for the PDA Detector
Verifying Operational Performance
Thermo Scientific
Surveyor PDA Plus Detector Hardware Manual
73
Figure 67.
Control page of the PDA Plus Diagnostics dialog box with discrete channels
settings
g. Click
Load To Detector
. A dialog box containing the message Method Has Been
Downloaded appears. Click
OK
.
h. Click the
Display
tab to open the Display page.
i.
In the Data area, click
Start
to refresh the display (see
).
j.
Save a printout or an electronic copy of the spectrum. Date the printout and add it to
your maintenance records. To monitor any degradation in light intensity, compare
this scan with similar scans that you obtain in the future (see
Note
These diodes roughly correspond to the wavelengths of 230 nm, 250 nm,
and 280 nm, respectively. The Discrete Channels area on the Display page
contains a digital readout for the selected diodes (see
). Record these
readout values to track the performance of the deuterium lamp. You can also use
other diode values to track the performance of the deuterium lamp.