ChromQuest Diagnostics for the PDA Detector
Calibrating the Detector with ChromQuest
66
Surveyor PDA Plus Detector Hardware Manual
Thermo Scientific
Figure 59.
Select Wavelength Calibration dialog box
c.
Click
Open
.
The holmium oxide bands of the selected file appear in the Wavelength File area as
shown in
.
Figure 60.
Diagnostics dialog box – Calibration page with the Holmium12 Wavelength
Calibration File selected
Note
ChromQuest has four calibration files to choose from. The HolmiumUV
file contains five wavelengths in the UV region while the other files, such as
Holmium12, use sets of wavelengths from both the UV and Visible wavelength
regions. The holmium oxide absorbance maxima are selected from a spectrum
published in “Holmium Oxide Solution Wavelength Standard from 240 to
640 nm - SRM 2034 (NIST Special Publication 260-54)”.