Xcalibur Diagnostics for the PDA Detector
Calibrating the PDA Detector
Thermo Scientific
Surveyor PDA Plus Detector Hardware Manual
35
Figure 29.
Page four of the Wavelength Calibration wizard
6. Rotate the Holmium Oxide filter wheel to position 2 as directed.
The Next button becomes active.
7. Click
Next
to proceed.
8. On page five of the Wavelength Calibration wizard, do the following:
a. Observe the status screen (see
).
The diagnostics program waits for a few seconds for the rise time filter to equilibrate
and takes a holmium oxide scan.
b. When the message
Click the Next button to proceed with calibration
appears
,
click
Next
to proceed.
Figure 30.
Page 5 of the Wavelength Calibration wizard
Note
The Next button is unavailable until the filter wheel is in position 2.