4
Xcalibur Diagnostics for the PDA Detector
Calibrating the PDA Detector
Thermo Scientific
Surveyor PDA Plus Detector Hardware Manual
41
Figure 39.
Page two of the Dark Current Calibration wizard
6. On page two (see
) of the Dark Current Calibration wizard, observe the status
readback as the calibration proceeds. When the calibration is complete, click
Next
.
Page three of the Dark Current Calibration wizard appears (see
). You can
export the results of the calibration from this page.
Figure 40.
Page three of the Dark Current Calibration wizard
7. To print a record of the Dark Current calibration:
a. Click
Export Results
.
The Save As dialog box appears (see
).
b. In the File name box, type a name. Then click
Save
.