ChromQuest Diagnostics for the PDA Detector
Calibrating the Detector with ChromQuest
64
Surveyor PDA Plus Detector Hardware Manual
Thermo Scientific
Preparing the Detector for Calibration
To prepare the Surveyor PDA Plus Detector for calibration
1. Replace the LC column with a flow restrictor. Then pump HPLC-grade water or
HPLC-grade methanol through the flow cell for at least one hour.
2. Turn on both lamps and wait one hour for the D2 lamp to equilibrate:
a. In the menu bar of the online Instrument window, choose
Control >
Instrument Status
to open the Instrument Status window.
b. Click the
Surveyor PDA Plus
tab to open the Surveyor PDA Plus page (see
c.
Under D2 Lamp, click
On
.
d. Under W Lamp, click
On
.
Figure 57.
Surveyor PDA Plus page of the ChromQuest Instrument Status window