Xcalibur Diagnostics for the PDA Detector
Verifying Operational Performance
Thermo Scientific
Surveyor PDA Plus Detector Hardware Manual
49
Viewing an Intensity Scan of the Deuterium Lamp
To view an intensity scan of the deuterium (D
2
) lamp
1. Create a display method to view the intensity of the deuterium lamp. Set channels A,B,
and C to monitor diodes 35, 52, and 77, respectively (see
).
See
“Creating a Display Method to View the Light Intensity”
for instructions
on creating a display method.
Figure 48.
Surveyor PDA Plus Method page with the selection of the Diodes/Intensity option
2. Load the display method to the detector as follows:
a. From the menu bar, choose
Surveyor PDA Plus > Direct Control
to open the
Direct Control dialog box.
b. Click the
Display
tab to open the Display page.
c.
In the Control area, click
Load Method
to open the Open Display Method
dialog box.
d. Browse to the display method that you created to monitor the deuterium lamp. Then
click
Open
to load the method.
The filename of the display method (.spda) appears in the box above the Spectrum
plot.
3. Click the
Configuration
tab to open the Configuration page. Then turn off the tungsten
lamp.