Configuration
Configuring the Surveyor PDA Plus Detector in ChromQuest
28
Surveyor PDA Plus Detector Hardware Manual
Thermo Scientific
Figure 23.
Detector Configuration dialog box
b. In the Detector Name box, type a name to identify the Surveyor PDA Plus Detector.
When you open the Instrument window, this name appears in the Analysis Channel
list.
c.
In the Detector Model list, select
PDA
.
d. In the Acquisition Source list, select
PDA Plus ACQ
.
e.
Click .
The PDA Plus Configuration dialog box appears.
Figure 24.
PDA Plus dialog box
f.
In the Stack boxes, type or select the appropriate stack address. Unless you are
controlling more than one Surveyor PDA Plus Detector from the same data system
computer, leave the value set to the default of 01.
The stack address must match the unit ID, which consists of two rotary switches, on
the back panel of the PDA detector.
g. Click
OK
to close the PDA Plus Configuration dialog box.
h. In the Detector Configuration dialog box, leave the Y-Axis units set to the default of
mAU and the Y-Axis Multiplier set to the default of 0.001.
i.
Click
OK
to close the Detector Configuration dialog box.