ChromQuest Diagnostics for the PDA Detector
Calibrating the Detector with ChromQuest
68
Surveyor PDA Plus Detector Hardware Manual
Thermo Scientific
8. In the Wavelength File area, check the delta values (see
a. If the delta values are not within the range of ±1 nm, repeat the wavelength
calibration procedure for verification.
b. If, after applying a new calibration, the delta values are still not within the range of
±1 nm, call your Thermo Fisher Scientific service representative for assistance.
Figure 64.
Diagnostics dialog box – Calibration page with a list of acceptable Delta values
9. If your data system computer is connected to a printer, click
to print a hard copy
report. To store the results, click
Export
.
The date and time of the wavelength calibration appear in the Wavelength area. This
information is also stored by the PDA detector.
Note
You can click
Cancel
in any of the Calibration dialog boxes at any time to abort the
calibration process.