ChromQuest Diagnostics for the PDA Detector
Calibrating the Detector with ChromQuest
Thermo Scientific
Surveyor PDA Plus Detector Hardware Manual
65
Performing a Wavelength Calibration
The alignment of the spectrum on the diode array is dependent upon the physical alignment
of various components of the optical bench. The alignment can become offset if the detector
is sharply jolted, for example, in shipping. Such bumps and jars can slightly change the
wavelength of light reaching the photodiode array. The automated wavelength calibration
determines the wavelength accuracy of the detector and uses the detector’s wavelength
algorithm to correct for any misalignment.
To perform a wavelength calibration with ChromQuest
1. On the Instrument Status page for the PDA detector, click
Diagnostics
to open the
PDA Plus Diagnostics dialog box.
2. Click the
Calibration
tab to display the Calibration page (see
Figure 58.
Calibration page of the PDA Plus Diagnostics dialog box
3. Open a Wavelength Calibration file as follows:
a. In the Wavelength File area, click
Open
to display the Select Wavelength Calibration
dialog box.
b. In the Select Wavelength Calibration dialog box shown in
, select an
appropriate wavelength calibration file from the list.
An appropriate wavelength file should include the range of wavelengths that you use
under normal operation conditions.