ChromQuest Diagnostics for the PDA Detector
Adjusting the Light Output from the Lamps
Thermo Scientific
Surveyor PDA Plus Detector Hardware Manual
81
c.
Click
Default
, and then verify that the following parameters are specified:
Start = 2
End = 511
Step = 1
Bandwidth = 1
Scan rate = 1
d. Click
Load To Detector
.
Determining the Diode of Maximum Intensity for the UV Range
To determine the diode of maximum intensity for the UV Range
1. Click the
Display
tab to open the Display page.
2. Click
Start,
which is
located in the upper right corner of the Display page.
3. Right-click in the top pane, which displays the intensity spectrum, to open the shortcut
menu.
4. From the shortcut menu, choose
Axis Setup
to open the Axis Properties dialog box (see
Figure 72.
Axis Properties dialog box