4
Xcalibur Diagnostics for the PDA Detector
Calibrating the PDA Detector
Thermo Scientific
Surveyor PDA Plus Detector Hardware Manual
37
11. Check that the delta values are within ±1 nm, and then do one of the following:
• If the Delta values are acceptable, proceed to
.
• If the delta values are not within the range of ±1 nm, do not export the results, and
proceed to
. Complete the calibration procedure, and then repeat the
wavelength calibration. If, after applying a second calibration, the Delta values are
still not within the range of ±1 nm, call your Thermo Fisher Scientific service
representative for assistance.
12. (Optional) Print a report of the calibration results:
a. Click
Export Results
to print the results to a file.
The Save As dialog box appears (see
).
Figure 34.
Save As dialog box
b. In the
Save As
dialog box, type a name, and then click
Save
.
Once you have saved the file with a name of your choice, you can view or print the
contents of the file using any text editing program (see